Replacement Procedures
8-4-7-1 Loading the System Software with USB memory stick(continued)
CAUTION
NOTE:
8-24
4. Press any key to continue when below message displays.
Figure 8-10. Upgrade message
5. Select one of the options for loading the system. Select
choice [a] to load the complete disk.
If you select [a], ALL existing software and data will be erased.
If backup has not been performed, all data like Patient
Database, System Configuration and User Configurations
(Customer Presets) will be lost.
•
To select [a], the complete disk will be loaded. This
option is recommended for application software
upgrade.
When to select [a] to load complete disk, please ensure
that any patient data on the disk has been backed up.
•
To select [b], only the bootable C: partition is loaded.
This option is intended for recovery of a system that will
not boot up. It is not recommended for application
software upgrade because during upgrade process, the
data on the system would possibly be impacted.
•
To select [c] to quit system upgrade process.
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10