Replacement Procedures
8-4-7-1 Loading the System Software with USB memory stick(continued)
CAUTION
NOTE:
8-24
4. Press any key to continue when below message displays.
Figure 8-10. Upgrade message
5. Select one of the options for loading the system. Select
choice [a] to load the complete disk.
If you select [a], ALL existing software and data will be erased.
If backup has not been performed, all data like Patient
Database, System Configuration and User Configurations
(Customer Presets) will be lost.
To select [a], the complete disk will be loaded. This
option is recommended for application software
upgrade.
When to select [a] to load complete disk, please ensure
that any patient data on the disk has been backed up.
To select [b], only the bootable C: partition is loaded.
This option is intended for recovery of a system that will
not boot up. It is not recommended for application
software upgrade because during upgrade process, the
data on the system would possibly be impacted.
To select [c] to quit system upgrade process.
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10