Site Preparations

2-2-3 Electrical requirements

2-2-3-1 General requirements
NOTE:
NOTE:
2-2-3-2 Electrical requirements for the Ultrasound system
2-6
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound system is only a conduit.
In the table below, the electrical specifications for the Ultrasound
system includes monitor and on board peripherals.
Table 2-3:
Electrical Specifications for Vivid T8/Vivid T8 Pro
Voltage
Tolerance
100-240 VAC
±10%
Vivid T8/Vivid T8 Pro – Basic Service Manual
system
Power Consumption
Max. 400VA
5718826-100 English Rev. 10
Frequency
50/ 60HZ