3-9-4 DICOM SR(continued)
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10
4. Select a DICOM storage devicce in the Selected devices
and then select Properties. The Properties window for the
selected DICOM storage device is displayed.
Figure 3-43. DICOM Storage Properties
5. Check the option Allow SR to enable DICOM SR.
Allow SR private data: send the current exam data ina
private format. This option is by default unchecked and
should only be used with DICOM storage devices that
can handle private fata format.
Signed Doppler velocities: send signed Doppler
veloities.
6. Select OK.
Connectivity setup
3-55