GE Vivid T8 Basic Service Manual page 148

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

System Setup
3-9-6-2 Create new dataflow with Worklist and DICOM Storage
3-64
1. Press F2 on the control panel.
2. Select Connectivity->Dataflow to display the dataflow
screen. Select Add to start the creation of a new dataflow.
Figure 3-50. Connectivity-> Dataflow
3. It brings up the Enter new name dialog. Enter a name, e.g.
Worklist-DICOM Storage, and then select OK.
Figure 3-51. Enter New Name
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10

Advertisement

Table of Contents
loading

Table of Contents