Care and Maintenance
10-7-5-2 Data Sheet for enclosure Source Leakage Current
Table 10-16:
Tester
Unit
Polarity
Power
Switch
Enter Name of tested peripheral here:
ON
NORM
ON
NORM
ON
REV
ON
REV
OFF
NORM
OFF
NORM
OFF
REV
OFF
REV
10-26
The test passes when all readings measure less than the value
shown in
Table 10-12 on page
Inspection Certificate.
Typical Data Sheet for enclosure Source Leakage Current
Tester
Neutral
or
Test 1
Ground
Speaker
Switch
Cover
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
10-21. Record all data on the PM
Test 2 Real
Panel Metal
Parts
Vivid T8/Vivid T8 Pro – Basic Service Manual