System Setup

Verify customer order

Physical inspection

EMI protection

3-18

Preparing for setup

Compare items received by the customer to that which is listed
on the delivery order. Report any items that are missing, back
ordered, or damaged.
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to 'Damage in
transportation' on page i-11 in the beginning of this manual.
The Vivid T8/ Vivid T8 Pro has been designed to minimize the
effects of Electro-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the
system from image artifacts caused by this interference. For this
reason, it is imperative that all covers and hardware are installed
and secured before the unit is put into operation.
See 'EMI limitations' on page 2-9 for more information about
EMI protection.
Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03