Optional Diagnostic Checks - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Care and Maintenance
Step
Item
10.
Power and
System Status
Indicators
11.
Battery

Optional Diagnostic Checks

View the Log
10-16
Table 10-10:
Physical checks (Continued)
Check for proper operation of all Power and System Status Indicators.
Where applicable, check that the battery is not damaged, does not leak,
does not emit an odor, and is not deformed or discolored. Observe all
warnings and cautions for battery handling, recharging, storing, and/or
disposal,
Optionally you can access the diagnostic software as described
in Chapter 5 or 7. View the error logs and run desired
diagnostics.
1. Review the system error log for any problems.
2. Check the temperature log to see if there are any trends that
could cause problems in the future.
3. Check the Configuration Log; update if needed.
Description
Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03