Care and Maintenance
Data Sheet for enclosure Source Leakage Current
Table 10-16:
Tester
Unit
Polarity
Power
Switch
Enter Name of tested peripheral here:
ON
NORM
ON
NORM
ON
REV
ON
REV
OFF
NORM
OFF
NORM
OFF
REV
OFF
REV
10-26
The test passes when all readings measure less than the value
shown in Table 10-12 on page 10-21. Record all data on the PM
Inspection Certificate.
Typical Data Sheet for enclosure Source Leakage Current
Tester
Neutral
or
Test 1
Ground
Speaker
Switch
Cover
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
Test 2 Real
Panel Metal
Parts