Setup connection to a DICOM server (continued)
CAUTION
NOTE:
NOTE:
Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
In Image Settings, setting compression to None may lead to
very long transfer time and loops larger than 500s frames will
be trucated. In case it happens, a warning is displayed on the
screen.
The pull-down menu gives these chioces:
Jpeg - JPEG/JPG compression (Default). JPEG is not
lossless.
None - (un-compressed data with huge files)
If you are using compression None, it is strongly
advised to reduce Frame Rate to 25 or 30 to avoid
truncating loops.
Rle - Run Length Encoding. This is lossless
compression.
Quality % - Set the wanted quality %. Default value is
95%.
If files are stored with a quality lower than 95% they will
occupy less space, but loose some quality.
Setting the quality % too low may add too much artifacts
to the images that they cannot be used for diagnosis.
If files are stored with a quality above 95%, the files will
grow much larger without almost no quality change.
Figure 3-32. DICOM SR Settings
Allow SR
If SR is enabled, all M&A that is supported by DICOM is
sent to the DICOM server (in a seperate job).
Allow SR Private Data
By default, this function is disabled. It is intended for
receivers that can handle "Private Data", such as
CA-1000.
Signed Doppler Velocities
It is disabled by default. It means that signs for Doppler
Velocities are removed in the DICOM SR.
Connectivity setup