GE Vivid T8 Basic Service Manual page 124

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

System Setup
NOTE:
3-48
Figure 3-35. DICOM Server Properties dialog
2. Select the "Smily" button to Ping the server.
By selecting the "Smily" button, a ping is sent to the remote
server to see if it is accessible via the network. It is not a
DICOM Echo (DICOM ping), so it does not check AE title or
port number.
Figure 3-36. Ping Check
3. If the network connection to the server is OK, it will be
illustrated by a smiling face
A sad smily indicates that the network connection is failing.
Typical causes:
Network cable not connected
Configuration erroe(s)
Vivid T8/ Vivid T8 Pro – Basic Service Manual
.
5490863-100 English Rev. 03

Advertisement