Quality Assurance - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Quality assurance

Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
In order to gain accreditation from organizations such as the
American College of Radiology (USA), it is the customer's
responsibility to have a quality assurance program in place for
each Ultrasound system. The program must be directed by a
medical physicists, the supervising radiologist/physician or
appropriate designee.
Routine quality control testing must occur regularly. The same
tests are performed during each period so that changes can be
monitored over time and effective corrective action can be
taken.
Testing results, corrective action and the effects of corrective
action must be documented and maintained on the site.
Your GE service representative can help you with establishing,
performing and maintaining records for a quality assurance
program. Contact GE for coverage and/or price for service.
Why do maintenance
10-5