Probes environmental requirements

Operation, storage and transport temperatures for probes
CAUTION
Temperature
Humidity
Pressure
CAUTION
CAUTION
NOTE:

Time and manpower requirements

Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
Probes should be operated, stored, or transported within the
parameters outlined below.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 2-6:
Probe Environmental Requirements
Operational
3° - 40°C
38° - 104°F
30 - 80%
non-condensing
700 - 1060hPa
Check the room temperature before you use the probe.
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Refer to Table 2-1 on page 2-3 to determine the needed
settlement time.
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
General Ultrasound system requirements
Storage
-5° - 50°C
23° - 122°F
10 - 90%
non-condensing
700 - 1060hPa
Transport
-5° - 50°C
23° - 122°F
10 - 90%
non-condensing
700 - 1060hPa
2-11