Connections on the I/O Rear Panel

NOTE:
Connect Ethernet
Connect USB Flash Drive
NOTE:

Connecting probes

Powering the system

Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC
standards (e.g. IEC60950 for data processing equipment and
IEC60601-1 for medical equipment). Furthermore, all complete
configurations shall comply with the valid version of the system
standard IEC60601-1. Everybody who connects additional
equipment to the signal input part or signal output part of Vivid
T8/ Vivid T8 Pro, configures a medical system, and is therefore
responsible that the Ultrasound system complies with the
requirements of the valid version of IEC60601-1. If in doubt,
consult the technical service department or your local
representative for GE Healthcare.
Connect the network cable to the Ethernet connector on the I/O
Rear Panel.
The connector is located on the rear side of Vivid T8/ Vivid T8
Pro.
USB Flash Drive approved for Vivid T8/ Vivid T8 Pro are verified
for EMC performance according to EN55011 class B. The use of
any other USB Flash Drive will compromise this verification, and
may cause interference on Vivid T8/ Vivid T8 Pro itself, or on
other electronic devices.
For approved models, please refer to Chapter 9.
Insert the USB Flash Drive in one of the USB ports on the Vivid
T8/ Vivid T8 Pro.
Please refer to User Manual on how to connect/disconnect a
probe.
Please refer to User Manual on how to power the system.
Completing the setup