DICOM SR

Vivid T8/ Vivid T8 Pro – Basic Service Manual
5490863-100 English Rev. 03
DICOM sTRUCTURED rEPORTING (SR) is a standardized for
medical results. Vivid T8/ Vivid T8 Pro supports the specialized
form for Adult Echo Ultrasound (TID 5200 Echocardiography
Procedure Report) and Vascular Ultrasound (TID 5100 Vascular
Ultrasound Procedure Report) for M&A results.
With the DICOM SR support, M&A for an exam can be sent at
the end of the exam or when exported from local archive. The
destination can be either a server on the network or a removable
media (DICOM Media), depending on the DICOM dataflow
selected.
TID5200 Echocardiography Procedure Report is senf if the
exam contains M&A from category Cardiac/Pediatric (heart) and
Vascular/Abdominal categories, two SR documents are sent.
TID5200 Echocardiography Procedure Report and TID 5100
Vascular Ultrasound Procedure Report do not support all M&A
results from Vivid T8/ Vivid T8 Pro. They are limited to the
following:
No unassigned measurement.
Not modifies Simpson or Bullet methods.
Basic derivations (Average, last, Min and Max), no
references between the derived measurements and the
ones they were made from.
Wall Motion Scoring: individual segment scores only
according to 16-segment model, no graded hypokinesis
(only Hypokinesis is used).
DICOM SR must be activated for each DICOM device.
1. Press F2 on the control panel and log on as admin.
2. Select Connectivity-> Dataflow, the dataflow sheet is
displayed.
3. Select the DICOM dataflow to configure in the Name
pull-down menu.
Connectivity setup
3-51