8.2.1 Calibration of temperature value
During temperature calibration only a check is carried out, i.e. no values are
changed.
1.
In the start display press the F1-key, the main menu will appear.
2.
Press
to select the menu item <6 - Temperature Test> and
acknowledge by
3.
Acknowledge by
4.
Press
to acknowledge the following instruction how to use a
temperature calibration set
5.
Select desired test temperature with
device heats up the set temperature.
6.
After 15 minutes the test is finished, an acoustic signal sounds. Compare the
test temperature with the displayed temperature of the DLB-A01N. If the two
values do not match, we recommend a temperature adjustment, see chap.
8.2.2
7.
Press
to cancel the calibration.