Calibration Of Temperature Value - KERN DLB 160-3A Operating Instructions Manual

Electronic moisture analyser
Hide thumbs Also See for DLB 160-3A:
Table of Contents

Advertisement

Available languages

Available languages

9.2.1 Calibration of temperature value

During temperature calibration only a check is carried out, i.e. no values are changed.
 Invoke service function „Tmp test", see chap.9.2
 Confirm by pressing the PRINT button. The current setting will be displayed. Use
the navigation buttons   to select the temperature during the check, available
35- 160°C, e.g. 120°C.
 Confirm using the PRINT button, the appliance heats up to the set temperature.
 After approx. 15 minutes compare the test temperature with the displayed
temperature of the DLB-A01N. If the two values do not match, we recommend a
temperature adjustment, see chap. 9.2.2
 Use the ON/OFF button to finish the calibration.
 Use the ON/OFF button to return into weighing mode.