15 Instant help
Possible causes of errors:
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault
Display is not lit up.
The display does not change
when a sample is being loaded • Sample tray / tray holder is fitted incorrectly.
The weight display changes
constantly / the stability display
does not appear.
Incorrect measuring result
Measurement is taking too
long
Measurement is
not reproducible
Drying does not start
TDLG_A-BA-e-2011
Possible cause
• The appliance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
•
Power supply interrupted.
•
Fuse has blown
• Sample tray has contact with wind protection
device or heated cover.
• Draught/air movement
• Table/floor vibrations
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• Check adjustment
• No resetting to zero before loading the sample
•
Incorrect shut-down criteria set
•
Sample is not homogenous
•
Drying time is too short
•
Drying temperature too high (e.g. oxidation
sample material, boiling point of sample
exceeded)
•
Temperature sensor soiled or defective
• Heated cover open
• The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
57