11.1.2 Adjust the switch-off criterion
A switch-off criterion defines under which conditions the device is supposed to cancel
the drying process. Cut-off criteria save continuous time checks and manual
cancelling of drying. Furthermore they ensure that measurements are always carried
out under the same conditions and provide for reproducible measurements. Two
different switch-off criteria can be selected; either automatic or time-controlled
In the start display press
The cursor displays the current setting
Press
to select the switch-off criterion and adjust by
Autostop
The integrated scale continuously determines the weight loss of
the sample during drying. If the loss of weight per time unit (30 s)
drops below the set nominal value, drying will be cancelled
automatically and the measurement result displayed.
Setting range:
Time
When this switch-off criterion is selected, the measurement will
continue until the set drying time has elapsed.
Setting range: 1 min – 99 min.
Acknowledge selection by
TDLG_A-BA-e-2011
, the current setting will be displayed;
• Auto (3mg/30s; factory setting)
• 1 mg/30 s – 10 mg/30 s, selectable
.
.
41