Tdlg 160-3-A; Calibration Of Temperature Value - KERN DLB A01N Operating Instructions Manual

Temperature calibrating set

Advertisement

2 TDLG 160-3-A

2.1 Calibration of temperature value

During temperature calibration only a check is carried out, i.e. no values are changed.
1.
In the start display press the F1-key, the main menu will appear.
2.
Press
3.
Acknowledge by
Press
calibration set DLB-A01N. The current setting is displayed
4.
Select desired test temperature with
The device heats up the set temperature.
5.
After 15 minutes the test is finished, an acoustic signal sounds. Compare the test
temperature with the displayed temperature of the DLB-A01N. If the two values
do not match, we recommend a temperature adjustment, see chap. 2.2
6.
Press
4
to select the menu item <6 - Temperature Test> .
.
to acknowledge the following instruction how to use a temperature
to cancel the calibration.
and acknowledge by
DLB_A01N-BA-e-2011
.