9.2.6 Auto zero

Under this menu item the automatic zero point correction can be switched on or off.
In switched-on-state the zero point is automatically corrected at drift or when dirty.
Note:
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the "stability
compensation". (e.g. slow flow of liquids from a container placed on the balance,
evaporating processes).
When apportioning involves small variations of weight, it is advisable to switch off this
function.
Press
to select the menu item <7 Auto zero> and acknowledge by
.
Press
to select the desired setting.
Options:
Auto zero off
Auto zero 1
Auto zero 2
Auto zero 3
Auto zero 3E
Acknowledge selection by
30
Auto Zero switched off
Auto Zero range ± ½ Digit
Auto Zero range ± 3 Digits
Auto Zero range ± 7 Digits
Auto Zero range ± 7 digits in the whole weighing range
.
TDLG_A-BA-e-2011