KERN & Sohn GmbH Ziegelei 1 Phone: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 E-Mail: info@kern-sohn.com Internet: www.kern-sohn.com Operating manual Electronic Moisture Analyser KERN DAB Version 1.1 10/2015 DAB-BA-e-1511...
Page 2
KERN DAB Version 1.1 10/2015 Operating manual Electronic Moisture Analyser Contents TECHNICAL DATA ..................... 5 APPLIANCE OVERVIEW ................... 7 2.1 Keyboard overview ........................9 2.2 Overview of display ........................10 BASIC INFORMATION (GENERAL) ..............11 3.1 Proper use ........................... 11 3.2 Improper Use ..........................
Page 3
Setting the drying period for time-controlled switching-off ........... 35 CARRYING OUT DRYING AND MEASURING PROCESSES ..... 36 RS 232 INTERFACE ..................39 11.1 Technical data ......................... 39 11.2 Setting the baud rate ......................39 11.3 Print out protocol: ........................40 DAB-BA-e-1511...
Page 4
Sample size / originally weighted in quantity ..............44 12.7 Drying temperature ........................ 45 12.8 Recommendations / Guidelines .................... 45 SERVICING, MAINTENANCE, DISPOSAL ..........46 13.1 Cleaning ..........................46 13.2 Servicing, maintenance ......................46 13.3 Disposal ........................... 46 INSTANT HELP ..................... 47 DECLARATION OF CONFORMITY .............. 48 DAB-BA-e-1511...
Page 5
Weighed-in quantity 10 g: 0.02 % Heating profiles Standard drying Fast drying Soft drying Linearity ± 0.003 g Stabilization time (typical) 3 sec. Warm-up time Recommended adjustment 100g (F2) weight, not added (class) 5°C….+40°C ambient temperature Environmental conditions max 80% air humidity non-condensing DAB-BA-e-1511...
Page 6
5 memory locations for measurement results Interface RS232 Dimensions (W x D x H) Housing 240 x 365 x 180 mm Available Ø 92 mm, 20 mm high drying room Net weight 4.8 kg Electric Supply 220V 50 Hz AC DAB-BA-e-1511...
Page 9
• Start / finish drying and measuring process • Taring • Confirm / store settings • Turn on/off • Select parameter (forward) • Increase digit • Switch on background illumination of display • Select parameter (backward) • Reduce digit • Switch off background illumination of display DAB-BA-e-1511...
Page 10
Designation Status display - drying process enabled Enabled heating profile flashes Standard drying Fast drying Soft drying Enabled switch-off criterion flashes Automatic switch-off Time controlled switch-off Manual switch off Previous drying time Current temperature Display in % moisture Subtotal DAB-BA-e-1511...
Page 11
Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. In KERN's accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
Page 12
4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. All language versions contain a non-binding translation. The original German is binding.
Page 13
4.2.1 Warning pictograms High voltage Chemical burn / corrosion General hazard Fire or explosion Poisoning Hot Surface 4.3 Personnel training The instrument may only be operated and maintained by trained personnel. DAB-BA-e-1511...
Page 14
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN instruments. Never make any modifications or design changes to the equipment whatsoever. Always use original spare parts and accessories.
Page 15
Substances that liberate corrosive gases when heated up (such as acids) In that case, work with smaller sample amounts as the liberated gases may condense on cooler casing parts and later cause corrosion. DAB-BA-e-1511...
Page 16
In this case, acclimatize the disconnected instrument for ca. 2 hours at room temperature. Avoid direct sunlight The air humidity should be between 45% and 75%, non-condensing. DAB-BA-e-1511...
Page 17
Take the moisture analyser carefully out of its packaging, remove the plastic jacket and install it at the designated work space. 6.2.1 Scope of delivery / serial accessories: • Moisture analyser, see illustration chap. 2 • 50 sample dishes • Power cable • Operating manual DAB-BA-e-1511...
Page 18
3. Place removal aid into the windshield. Ensure that the lug lies exactly in the groove of the windshield. 4. Attach the sample dish holder carefully and rotate it until it locks. 5. Put the sample dish on the removal aid. DAB-BA-e-1511...
Page 19
Precise alignment and stable installation are a precondition for repeatable results. To compensate for minor unevenness or tilts of the base, level the instrument. Level instrument with foot screws until the air bubble of the water balance is in the prescribed circle. Check levelling regularly DAB-BA-e-1511...
Page 20
If matching, connect the moisture analyser. 6.6 Switching on and off Turn on appliance by pressing the ON/OFF key. The display lights up and the equipment conducts a self-test. Wait until basic configuration appears. To switch off press the ON/OFF button, the display disappears. DAB-BA-e-1511...
Page 21
6.7 Connection of peripheral devices Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the moisture analyser from the power supply. Only use accessories and peripheral devices by KERN, as they are ideally tuned to the appliance. DAB-BA-e-1511...
Page 22
• Carry out adjustment as near as possible to the balance’s maximum weight (recommended adjustment weight see chap. 1). Info about test weights can be found on the Internet at: http://www.kern-sohn.com. Press and keep pressed until „CAL“ is displayed.
Page 23
Put the required adjustment weight carefully into the centre of the sample dish and close heating hood. Adjustment is carried out automatically. After successful adjustment the balance automatically returns to weighing mode, „100,000 g“ will be displayed. Take away adjustment weight. Wait until „0.000 g“ appears. DAB-BA-e-1511...
Page 24
Temperature calibration We recommend performing occasional tests of the instrument’s temperature value, using the optional temperature calibration set KERN DAB-A01. Before you do this, allow the device to cool down for at least 3 hours after the last heating phase.
Page 25
<1> and confirm by . The currently set heating profile flashes. Using arrow keys select heating profile <SOFT> and confirm by The currently set heating profile and switch-off criterion flashes. Use the arrow keys to select switch-off criterion <MANUAL>. DAB-BA-e-1511...
Page 26
Use the arrow keys to enter the desired test temperature. For the first or the first two digits select a value and confirm by Using arrow keys select the desired value for the last digit. Confirm input by DAB-BA-e-1511...
Page 27
„8.2X“ will be displayed. If not, separate device from power supply and restart step two. 3. Adjust temperature using arrow keys . Each time you press the button the temperature will be increased or reduced by 1°C. 4. Confirm input by DAB-BA-e-1511...
Page 28
The current temperature, lapsed time as well as the current interim result will be continuously updated and displayed. The icons of the enabled heating profile and the switch-off criterion are flashing. Using the display con be switched over to the different displays of results. DAB-BA-e-1511...
Page 29
Open the heating hood and remove the sample with the help of the removal aid. Caution: Sample dish and all parts of the sample chamber are hot! Using the result in the display will be deleted. The display returns to weighing mode. The device is now ready for another measurement. DAB-BA-e-1511...
Page 30
This memory location is provided for the factory settings (standard drying 120°C, automatic switch-off) and cannot be changed. Use the arrow keys to select the desired ident number and confirm by . The currently set heating profile flashes. DAB-BA-e-1511...
Page 31
Options: Standard drying Fast drying Soft drying Standard drying <STANDARD> This heating profile is suitable for most samples. The sample will be heated drying temperature will kept constantly this temperature. Selectable 40°C -199°C DAB-BA-e-1511...
Page 32
(such as easily liquefiable substances or substances containing sugar). Skin formation affects evaporation of trapped moisture. The temperature will increased continuously and will not reach the set drying temperature before the so-called ramp duration has elapsed. DAB-BA-e-1511...
Page 33
When this switch-off criterion is selected the measurement will continue until the set Time-controlled switch off, drying time has lapsed (max. time 99h 59 selectable 3 min – 99h 59 min min). (10s steps) Manual switch-off by pressing stop key <MANUAL> DAB-BA-e-1511...
Page 34
All settings are stored under the ident number <1> until a new method is allocated to the ident number. <AUTO> <MANUAL> When <TIMING> is selected, the display of setting the drying period of time is flashing, see step „5“. DAB-BA-e-1511...
Page 35
Confirm input by . All parameter settings will be applied to a subsequent drying process and displayed on screen. All settings are stored under the ident number <1> until a new method is allocated to the ident number. DAB-BA-e-1511...
Page 36
After pressing the button the first ident number <0> of a drying method will be displayed. Use arrow keys to select a desired ident number of a drying method and confirm by . The currently set heating profile flashes. DAB-BA-e-1511...
Page 37
The current temperature, lapsed time as well as the current interim result will be continuously updated and displayed. The icons of the enabled heating profile and the switch-off criterion are flashing. Using the display con be switched over to the different displays of results. DAB-BA-e-1511...
Page 38
Open the heating hood and remove the sample with the help of the removal aid. Caution: Sample dish and all parts of the sample chamber are hot! Using the result in the display will be deleted. The display returns to weighing mode. The device is now ready for another measurement. DAB-BA-e-1511...
Page 39
Disconnect moisture analyser from the power supply and connect to the appliance interface with a suitable cable. Faultless operation requires an adequate KERN interface cable. Communication parameters of the RS 232 interface of moisture analyser and printer must match.
Page 40
„1“. The instrument stores the last five protocols (ident number 1 -5). With every new measurement, the ident number One will be overwritten. Use the arrow keys to select the desired ident number and confirm by . The protocol will be edited. DAB-BA-e-1511...
Page 41
Sample protocol (KERN YKB-01N): MOISTURE DETERMINATION (1) Protocol ident number (1) Gerätetyp Machine Type: DAB 100-3 Heating Mode: STANDARD Selected heating profile Stop Mode: AUTO STOP Selected switch-off criterion Drying temperature Heating Temp: 120 °C Time elapsed: 03:52 Overall duration of drying process Wet W: 10.145 g...
Page 42
The radiation applied in the KERN DAB penetrates mainly the sample in order to be transformed inside it into heat energy that is, warming from the inside to the outside. A minor amount of radiation is reflected by the sample, a reflection that is less in dark samples than in light-coloured ones.
Page 43
For liquids, pastes or melting samples we recommend to use a glass fiber filter. The glass fiber filter has the following advantages: • even distribution thanks to capillary attraction • no formation of droplets • fast evaporation due to a greater surface DAB-BA-e-1511...
Page 44
In the course of this arise two opposed requirements: The lighter the originally weighted in quantity, the easier it is to achieve shorter drying times. However, the heavier the originally weighted in quantity, the more accurate a result. DAB-BA-e-1511...
Page 45
The glass fibre filter is also useful as a protection when splashing materials are dealt with (each splash falsifies the final result). You will find examples taken from actual use in our application manual, available from the KERN home page (www.kern-sohn.com). DAB-BA-e-1511...
Page 46
Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner. 13.2 Servicing, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Ensure that the balance is regularly calibrated, see chap. Monitoring of test resources. 13.3 Disposal ...
Page 47
• Drying temperature too high (e.g. oxidation sample material, boiling point of sample exceeded) • Temperature sensor soiled or defective • Heating hood open Drying does not start • The mains supply connection has been interrupted (mains cable not plugged in/faulty). DAB-BA-e-1511...
Page 48
EN 50581 :2012 Date 29.07.2015 Signatur Date Signature Ort der Ausstellung 72336 Balingen Albert Sauter Place of issue KERN & Sohn GmbH Manager Managing director KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com DAB-BA-e-1511...