8.2 Calibrate / adjust temperature of the heating module

8.2.1

Temperature calibration

We recommend performing occasional tests of the instrument's temperature value,
using the optional temperature calibration set KERN DAB-A01.
Before you do this, allow the device to cool down for at least 3 hours after the last
heating phase.
Preparation:
 Switch off moisture analyser
 Install temperature calibration set as per illustration
 Turn on moisture analyser