8.2.2 Adjust temperature
If in temperature calibration the admitted divergence is exceed or not reached, the
temperature of the instrument can be adjusted as described below.
1.
Select menu item <7 – Heater Calibration> (see chap. 8.2.1, step 1) and
confirm by
2.
Confirm the inquiries <Are you sure> and the following instruction how to use
a temperature calibration set by
The current settings of the both temperature points T-lo / T-hi are displayed.
3.
Press
select between <T-lo> and <T-hi>. Acknowledge with
4.
Press
, the first heating phase is started.
5.
Temperature calibration for the first point takes 15 min. Compare the
temperature displayed on the DLB-A01N with that of the moisture analyzer.
If the two values do not match, correct with
6.
Confirm by