6 Calibration Procedure

1. Switch-off the device.
2. While pressing the
3 times and releases the
count (e.g. 1933).
3. Press the
4. Load calibration weight (40kg). The display shows the internal span count
(e.g. 1146).
5. Press the
the calibration weight (40.0 kg) appears.
6. Remove the calibration weight and switch-off the device.
Calibration is completed.
MODEL
MAP 80K1
8
/zero key, after 1 second hit the Mode key sequentially
/zero key. The display shows the internal zero
/zero key once. The display appears [
/zero key once to start the calibration. Shortly after the value of
ZERO Count
500-3000
0 ].
Calibration
SPAN Count
500-2700
Weight
40kg
MAP-SH-e-1010