7.6 Temperature calibration / adjustment

The temperature can be checked and re-adjusted if necessary using the optional
temperature calibration set (MLB-A03).
Temperature calibration
The balance is in normal weighing mode.
-
Insert the temperature sensor through the balance cover opening and push in
-
close to the weighing plate (see pictures).
-
-
Press the
key and then press the
-
A "Code" is now queried on the display screen.
Use the arrow keys to enter the correct code ("002003").
-
Numerical entry from 0-9
For changing the entry point on the display screen
Press the
key. "St_tE " will appear on the display screen - the temperature
-
sensor setting programme.
Press the
key. The temperature inside the weighing region will now appear.
-
If this is not 20°C, please use the arrow keys to alter the temperature.
Press the
key. If a temperature of 20°C is entered, a number of approxi-
-
mately 4000 will appear on the display screen.
The
key is now used to start temperature calibration. The balance will now
-
heat up to 50% of its total power.
MLB50-3-BA-defsi-0310
Guide bush for temperature sensor
and
keys simultaneously.