10.1.2 Connect the preheating stage

If necessary, the function „Preheating" can be activated for preheating the sample
chamber before starting the measurement.
A preheated device may influence the reproducibility of the results as the device is in
the same temperature status at every measurement.
Press
to select the menu item <2 Preheating> and acknowledge by
.
Mit Press
.
Procedure with activated function:
In the start display press the button under
instructions. Close the lid
Wait until the set temperature is reached.
When the <Ready> display appears, press the button under
Put the empty sample tray in the sample chamber. Close the lid, the
scale will automatically be reset to zero.
Weigh-in the prepared sample and close the lid. Measurement will be started
automatically.
For further steps, see chap. 11.2.
For standard applications, a preheating normally is not required.
36
to switch on or off the preheating stage and confirm by
and follow the user-guide
.