1 Technical data Data DLT 100-3 Radiator Halogen (1 x 400 W) 35°C - 160°C Temperature range Choice of steps at 1°C Maximum load (Max) 160 g Warm-up time Minimum for drying 0.5 g Weighing mode 0.001g Readability (d) Moisture analysis mode 0.01 % Weighing mode 0.001g...
Page 6
Shutoff criterion • Manual Press stop button • Time Drying is finished after the set time, adjustable 1 – 99 minutes • Autostop % when the set weight loss (%) per time unit is lower than the nominal value (both values adjustable) •...
2.2 Tablet Front side: Rear side: 1 Touch screen 1 Audio loudspeaker 2 Front camera 2 Volume key 3 On / off key 4 Loudspeaker / headset port 5 Microphone 6 Micro USB 2.0 port 7 Rear camera 8 MicroSD card slot DLT_N-BA-e-1421...
Information is available on KERN’s home page (www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. In KERN's accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances. All language versions contain a non-binding translation. The original German is binding.
Page 11
CAUTION The moisture analyzer operates using heat! Maintain sufficient space in the environment of the instrument to prevent heat build-up (distance from the instrument 20cm, upwards 1m). The heat extractor of the sample must never be covered, blocked, taped up or altered in any other way. Never place combustible materials on, under or next to the instrument, as the environment of the instrument heats up to a high temperature.
5 Transport and storage 5.1 Testing upon acceptance When receiving the appliance, please check packaging immediately, and the instrument itself when unpacking for possible visible damage. 5.2 Packaging / return transport Keep all parts of the original packaging for a possibly required return.
Page 13
On the installation site observe the following: Remove explosion prone and easily flammable material in the immediate vicinity. Emerging vapours, sample dish and all parts of the sample chamber are hot! Protect the instrument against direct draughts due to open windows and doors.
6.2 Unpacking and placing 1. Drying unit Take the instrument carefully out of its packaging, remove the plastic jacket and install it at the designated work space. The instrument is supplied part-assembled. Immediately after unpacking check if the delivered items are complete. Assemble the separate component parts according to their sequence.
5. Level instrument with foot screws until the air bubble of the water balance is in the prescribed circle. Check levelling regularly. 5. Tablet Remove the tablet carefully from the packaging, remove plastic cover and install at intended place of installation, using the table mount. Charge the rechargeable battery.
6.5 Mains connection Power is supplied to the integrated balance via the external power unit. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.
Page 20
If not, activate it by following the steps below: In weighing mode, press and hold pressed until the audio signal stops. Release the key and the menu item “units“ will appear. Press repeatedly until “tbL mode“ appears. ...
Page 21
The start screen will appear. Ensuring Bluetooth connection When enabled, the icon for Bluetooth will be shown in the top right corner. To start application “moisture analysis“, use DLT_N-BA-e-1421...
Page 22
Wait for Bluetooth connection, whereupon the start screen will appear. Setting operator language On delivery the display is set to German. Setting additional languages: Tap <Settings>. Tap <Language & input>. Tap <Language>. DLT_N-BA-e-1421...
Page 23
In dropdown menu select the desired language. Initial Commissioning In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1). During this warming up time the weighing system must be connected to the power supply (mains, accumulator or battery).
6.8 Connection of peripheral devices Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the moisture analyser from the power supply. Only use accessories and peripheral instruments by KERN, as they are ideally tuned to the instrument. DLT_N-BA-e-1421...
Adjustment in adjustment mode “external user define“ can also be carried out, based on weights of different nominal values, however, from a metrological point of view this does not produce best results. • Info about test weights can be found on the Internet at: http://www.kern- sohn.com. Call function: ...
Page 26
During the flashing display put the required adjustment weight carefully in the centre of the sample dish. After successful adjustment the balance automatically returns to weighing mode. Take away adjustment weight To view/issue an adjustment protocol, call the following menu item (See chap 10.5 ): <Settings>...
7 Operating elements In “tablet mode” the keys of the drying unit will be disabled (with the exception of ON/OFF). Use the tablet for all actions. 7.1 Start screen DLT_N-BA-e-1421...
Page 28
Call-up function menu: Status display Tap and hold: Set standby temperature: User specific shortcut keys (M1- M5) for calling up the 5 most frequently used drying programs Function keys (Pre tare, zero setting / tare, print) Link to android service programs ...
7.2 Work screen The work screen is shown at the start of measuring, see chap. 11. Drying process active Instructions for the performance of tasks during measuring Weighing guide; only shown when function is enabled, see chap. 11.1.4 Enter tare value numerically Start drying;...
8 Simple weighing / taring In order to obtain exact weighing results, the integrated balance must have reached its operating temperature (see warming up time chap. 1). 8.1 Weighing: Ensure weighing mode Wait for zero display, reset to zero using 0/T.
9 User profile In this menu 10 user profiles can be created, processed or deleted. Each user will be assigned to a profile with his own specific settings (basic settings of the instrument, hole data memory). Factory setting is „Default“, this cannot be deleted or renamed. ...
10 Setup menu In this menu item the user-specific basic settings of the instrument take place. The settings are stored together with the active user profile and remain valid if working with that profile. If a user profile is called up, the respective settings are automatically loaded, see chap.
10.1 Heating module settings Tap <Heater parameters>, see chap.: 10. The available settings will be displayed. Settings: <Standby-temperature> Switched off Standby The numeric input window will be Temperature displayed. Enter required temperature selectable 35°C – 100°C. To import press <OK>. The display returns to the previous screen.
Page 34
< Stability test > Settings: Stability control before start of measurement switched off Stability control before start of measurement switched off < End acoustic signal> Settings: Switched off When turned on, an audio signal will be sounded as soon as the test is finished.
10.2 Balance settings Tap <Weighing parameters>, see chap: 10. The available settings will be displayed. Settings: < Filter > Filter 1 Setting for dosage This menu item allows the balance to be set according to Filter 2 The balance reacts quickly and in a specific ambient conditions and sensitive manner, very quiet set-up measuring purposes.
Tap <Peripheral>, see chap: 10. The available settings will be displayed. Settings: < Printer selection > DPP-250 Setting for KERN YKC-01 DPP-350 < Print at stability > Settings: Stability check before data transfer on Stability check before data transfer off Settings: <Print heading>...
10.4 Adjustment Tap < Calibration settings > see, chap: 10. The available settings will be displayed. < Calibration data > View or print adjustment protocol: < Calibration mode > Settings: External Here the adjustment takes place near the default maximum load of the balance Standard (recommended adjustment weight see...
10.6 Service menu Tap <Service >, see chap: 10. The available settings will be displayed. The changes to service menu settings described below must be made by expert staff with profound knowledge of balances. < Backup and reset system data > <...
You will find examples taken from actual use in our application manual, available from our KERN home page (www.kern-sohn.com). The instrument offers the possibility to set the drying parameters sample-specific manually or to load a suitable drying program with the required drying parameters from the database.
11.1.1 Setting the heating profile This function offers different heating profiles to adapt the drying characteristic optimally to the used sample. in dropdown menu < Heating profile > (see chap. 11.1). The available heating profiles are displayed: • Standard • Fast •...
Page 42
3. Gentle drying This heating profile is suitable for the soft drying of substances, which tend to skin formation (e.g. substances containing Display pictogram during drying sugar or volatile substances). Skin formation will then affect the evaporation of the trapped moisture. The temperature is permanently increased and reaches the selected drying temperature only after finishing the so called ramp duration.
Page 43
4. Step-by-step drying This heating profile is suitable for drying substances which consist of several components (such as etheric oils), which Display pictogram during drying evaporate at different temperatures or of substances which present a specific behaviour when being warmed up. 3 steps can be selected.
11.1.2 Shutoff criterion A switch-off criterion defines when the instrument has to finish the drying process. The switch-off criteria make sure that the measurements are always finished under the same conditions thus providing repeatable measurements. Tap< Shutoff Criterion > in dropdown menu (see chap.
Page 45
Selectable parameters: Variation: 0.1 - 10.0% Time unit: 1 – 99 s To import press <OK>. This switch-off criterion is based of the Absolute auto stop weight loss (mg) per time unit. Measurement is finished, when the set weight value per time unit is lower than the nominal value, both values being freely adjustable [mg / sec].
11.1.3 Preheating (before weighing-in) If needed, the function „preheating“ can be enabled for preheating the sample chamber prior to beginning the actual measurement. A preheated device may affect the consistency of the results as the device keeps the same temperature status for each measurement. Preheating will take place up to the point of measuring, in standby mode (For setting see chap.
11.1.4 Weighing assistance The weigh-in aid facilitates the weighing-in of the samples on a nominal value with adjustable tolerance range. This is especially useful, when samples with the same weight are processed. This will increase reproducibility of the measurement results. If all samples are weighed-in within the tolerance range, this will increase repeatability.
11.1.5 Result display This function will define the unity where the measuring results are displayed and printed out. Calculation: Display shows the moisture content of the sample as a % M: Moisture content percentage of the wet weight (W = wet weight = starting weight = 100%) Wet weight W –...
Page 50
Settings: Tap < Units > in dropdown menu (See chap. 11.1). Selectable display types: Moisture (%M) Dry weight (%R) Atro (%A) Residual weight (g) Note: You can switch over during or after weighing to the available units, see chap. 12 “Switch over result display“.
11.2 Save drying parameters in database Tap < Moisture Determination >. Tap <Database parameters setup>. The dropdown menu for data records will appear. The database will be empty when started up for the first time. To create a new data record (drying program), tap the icon in the top right corner.
11.2.1 Creating a new drying program (data record): Call parameter dropdown menu, see chap. 11.2 Tap <Record name> and the input window will appear. Enter name (such as maize silage) for the new drying program (max. 30 characters) and import by pressing <OK>.
11.2.2 Change / delete drying programs (data record) Tap <Moisture Determination>. Tap <Database parameters setup>. The dropdown menu for data records will appear. Highlight the data record to be edited by tapping on it. To highlight all data records, tap icon and select <...
12 Carrying Out Measurement During the first start-up set the drying parameters as described in chap. 11. (1) Start-up Turning on moisture meter, see chap. 6.6. In order to obtain exact results, the instrument must have reached its operating temperature (see warm-up time chap.
Page 55
<Manual parameter setting> After setting all drying parameters, tap the control button the top right corner. The work screen will appear, showing the previously set parameters. < Database parameter settings > Load desired drying program from the dropdown menu by tapping. to search names in the dropdown menu.
Page 56
Shortcut keys Tap shortcut key (M1 –M5). The work screen for the selected drying program will appear. DLT_N-BA-e-1421...
Page 57
(3) Carrying Out Measurement Step 1: Preheat Work screen without preheating, Work screen with preheating, see chap.11.1.3 see chap.11.1.3 Sample display: Work screen with weighing The display may prompt you to close the sample guide disabled, see chap: 11.1.4 chamber.
Page 58
Working with the weigh-in aid: The bar graph display moves from the left to the right and proceeds equally to the weight loaded onto the weighing balance. yellow: Weighed-in < lower weight value green Target weight within the weight limits red: Weighed-in >...
Page 59
Step 4: Display during drying process You can watch the drying process on the display unit. The drying progress is shown continuously as a graphic display. The display shows the selected settings. To interrupt the drying process tap control button Display of results, tap for switching over Current measured value in % Current temperature...
Page 60
Switch over display of result Tap result display of result and import desired unit into dropdown menu (such as [g] ). Switching in graphic display The display may be zoomed in or out by tapping the diagram. DLT_N-BA-e-1421...
Page 61
Step 6: End of measurement, display/printout measuring results The end of the drying process will be indicated by an audio signal, provided this has been enabled in the menu (menu settings <end audio signal “ON”>), see chap. 10.1) whereupon heating will be turned off. ...
Page 63
Step 7: Save measurement results (max. 300 measurement results) Tap control button Enter data record name and designation and import by pressing The database results will be displayed. DLT_N-BA-e-1421...
13 Call up / edit / delete measurement results 13.1 Calling measurement results Tap <results>. Tap <database results>. The list of data records will be displayed. Select data record. The selected measurement result will be displayed. DLT_N-BA-e-1421...
13.2 Changing / deleting measurement results Tap <results>. Tap <database results>. The list of data records will be displayed. Highlight the data record to be edited by tapping To highlight all data records, tap icon and select < Select all > ...
14 GLP Tap <GLP settings>. Select desired setting. 1. <Manual setting> Manual setup < >. Selectable parameters: Id Balance Id Project Id User Tap the respective control button, the numeric input window will appear. Import entry with , the display returns to <OK>...
Page 67
Tap <Record name> and the input window will appear. Enter designation (max. 30 characters) and import by pressing <OK>. Then enter all the other parameters: Id Balance Id Project Id User To save new data record to database, press The list of data records will be displayed.
The radiation applied in the KERN DLT penetrates mainly the sample in order to be transformed inside it into heat energy that is, warming from the inside to the outside. A minor amount of radiation is reflected by the sample, a reflection that is less in dark samples than in light-coloured ones.
15.4 Preparing a sample Prepare one sample at a time for measuring. This prevents the sample from exchanging moisture with its surroundings. If several samples have to be taken at the same time, they should be packed in airtight boxes so that they do not undergo changes during storage.
15.5 Sample material Easy to determine are usually samples with the following characteristics: Grainy to powdery, pourable solids Thermally stable materials, emitting the moisture to be determined easily without other substances evaporating at the same time Liquids that vaporize to leave a dry substance without developing a film Difficult to determine may be samples that are: ...
15.7 Drying temperature Bear in mind the following factors when setting the drying temperature: 15.7.1 Surface of the sample: Compared with powdery or grainy samples, liquid and spreadable samples have a smaller surface for the transmission of heat energy. The use of a glass fibre filter improves the heat application. 15.7.2 Colour of sample: Light-coloured samples reflect more heat radiation than dark ones and therefore require a higher drying temperature.
15.8.3 Table of applications: Drying Solid Weight Drying Moisture period body Material Sample temperature (approx.) (°C) (approx.) (approx.) (min) ABS (Novodur P2H-AT) 0.11 Accumulator lead 0.19 Acryl granulate 10-15 0.18 Activated carbon 13.33 Activated carbon 6.12 Sliced pineapple 14.4 6.71 Sliced apple (dry) 10-15 76.5...
Page 73
Drying Solid Weight Drying Moisture period body Material Sample temperature (approx.) ( °C) (approx.) (min) (approx.) Setting lotion 0.01 98.76 Setting lotion (extra strong) 97.85 Hair styling gel 37.0 94.71 Oat flakes 9.35 Hazelnut cores Hazelnut cores (peeled) 3.74 Hydranal sodium tatrate – 2 15.67 –...
Page 74
Drying Solid Weight Drying Moisture period body Material Sample temperature (approx.) ( °C) (approx.) (approx.) (min) Paper PA 6 (Ultramide B3WG5) 0.05 PA 6.6 (Ultramide B3WG5) 0.15 PBTP (Crastin SK645FR) 0.05 PC (Macrolon 2805) 10-12 0.08 PC/ABS (Babyblend T65MN) 9-11 0.12 Pepper, black, powder 7.97...
Page 75
7.32 Wheat spring water Sausage casing 78.56 Toothpaste 34.28 Pulp 7.32 Cement 8-12 Sugar 11.9 Sugar beets 13.4 30.94 You will find further examples taken from actual use in our application manual, available from the KERN home page (www.kern-sohn.com). DLT_N-BA-e-1421...
Remove any spilt sample material immediately. 16.2 Servicing, maintenance The instrument may only be opened by trained service technicians who are authorized by KERN. Ensure that the integrated balance is regularly calibrated, see chap. Monitoring of test resources. 16.3 Disposal...
17 Instant help Fault Possible cause Display is not lit up. • The instrument is not switched on. • The rechargeable battery is empty. The display does not change when a sample is being loaded • Sample dish / dish holder is fitted incorrectly. •...
Need help?
Do you have a question about the DLT-N and is the answer not in the manual?
Questions and answers