Repair
The NBP
Remove the NBP Board either to replace it or to access the Inverter Board or the Display. The
Board
Main Board can have been removed or it can be in place.
Removal of
Step 1.
NBP Board
Step 2.
Installation of
The Display and Inverter Board with shield must be in place and must be connected. The
NBP Board
Inverter-to-Main and Display-to-Main Cables will run under the NBP Board and must be in
place now.
Step 1.
Step 2.
Step 3.
5-38
Troubleshooting, Service and Repair
Figure 5-9 Removal of the NBP Board
If necessary, press the spring tab on the NBP-to-Main Connector and disconnect the
NBP-to-Main Cable from the Main Board.
Note—
If the Main Board has been removed from the Chassis, this cable may already
have been disconnected.
Remove the four 6mm T-10 screws holding the NPB Board to the Chassis and lift
the board from the Chassis.
Place the NBP Board on the Chassis with the pump facing up and on the outside
edge. The holes in the board should match the mounts on the chassis. Be sure the
two cables are under the board, and that one screw mount divides the ribbon cable.
Secure the board to the Chassis with four 6mm T-10 screws.
Make sure the NBP-to-Main Cable is attached to the connector on the NBP Board.
6 mm T-10
NBP-to-Main Cable
Connectors