Nbp Measurement Specifications For Telemon B - Philips TeleMon B M2636B Service Manual

Table of Contents

Advertisement

NBP Measurement
Specifications for
TeleMon B
Shipping & Storage
Temperature Range
Altitude Range
Humidity Range
For ECG, ST, and %SpO
Note—
System, Series C, User's Guide.
Measurement Range
Initial Inflation Pressure
Maximum Pressure
Overpressure Safety Limit
Accuracy
Heart Rate Range
Cuff Deflation Rate
NBP Alarm range at Central 10 to 260 mmHg
Rated Life
Cuff Inflation Time
Limitations on Use
Auto Mode Repetition Time 5,10, 15, 30, 60 or 120 minutes
Typical Measurement Time
@ HR > 60 b/min in Auto
or Manual Mode
0º C to 60º C (32º F to 140º F) / without battery
0 to 50º C (32 to 122º F) / with battery
Up to 3048 m (10,000 ft.)
< 90% relative humidity, non-condensing
specifications, please refer to the Telemetry
2
Systolic: 40 to 260 mmHg
Diastolic: 25 to 200 mmHg
Adult: 160 mmHg
Pediatric: 120 mmHg
280 mmHg
300 mmHg (max)
+/- 3 mmHg or 5% of reading (whichever is
greater)
40 to 200 b/min
6 mmHg/step
30,000 measurement cycles (36/day for 2.3
years)
60 sec man (adult or pediatric)
Not for use on neonates.
When used on patients suffering from cardiac
arrhythmias, information presented will have
undefined accuracy.
Average: 30 seconds
Maximum: 170 seconds