Philips TeleMon B M2636B Service Manual page 85

Table of Contents

Advertisement

Test Block
Test or "Inspection" to Perform
Name
Performance
Accuracy Test
Test NBP:
Step Action
1
2
3
4
5
6
Leakage Test
Step Action
1
2
Connect the manometer and the
pump to the NBP connector.
Go to Service Mode.
Go to the Calibration
application.
Raise the pressure to 280
mmHg with the manometer and
pump.
Wait 10 seconds for the
measurement to stabilize.
Compare the manometer values
with the displayed values.
Document the value displayed
by the Monitor. If the
difference is greater than
3mmHg then calibrate the
module.
Watch the pressure value for 60
seconds. After 60 seconds the
value should have decreased by
less than 6 mmHg.
Calculate and document:
Leakage test = Accuracy -
displayed value.
M2600A Telemetry System - Test and Inspection Matrix
Expected Test Results
Value displayed on Monitor =
x1
If difference <=3mm proceed
to next test
Leakage test value = x2
If <= 6mmHg proceed to next
test
Test and Inspection Matrix
What to Record
on Service
Record
6-3