M2600A Telemetry System - Test and Inspection Matrix

M2600A Telemetry System - Test and Inspection Matrix
Test Block
Test or "Inspection" to Perform
Name
Visual Test:
Inspect the system (and packing material if
applicable) for obvious signs of damage.
Also check external leads and accessories.
Power On:
Step Action
1
2
3
6-2
Test and Inspection Matrix
Switch on the M2636B.
Observe that the monitor boots
up successfully without
displaying any error codes and
that output appears on the
display (example waveform
and parameter labels). An
indication that the "battery
needs reconditioning" is
acceptable.
Insert a functioning Telemetry
transmitter that is compatible
with the M2636B. Observe that
the "Transmitter Disconnected"
prompt disappears when the
transmitter is docked.
Expected Test Results
The system does not have any
obvious signs of damage =
Pass.
Monitor boots up displaying
no error codes, responds
correctly to telemetry
transmitter = Pass
What to Record
on Service
Record
V:P or V:F
where P=Pass and
F=Fail
PO:P or PO:F
where P = Pass
and F = Fail

Advertisement