Step 4.
The Inverter
Remove the Inverter Board to replace it or to access the Display. Before removing the
Board
Inverter Board, remove the NBP Board. The Main Board can have been removed or can be in
place.
Removal of
Step 1.
Inverter Board
If the Main Board is in place, press the connector of the NBP-to-Main Cable into
the keyed connector on the Main Board. If the Main Board is not in place, perform
this step when you mount the Main Board.
Inverter Shield
IR Shield
Figure 5-10 The Inverter and IR Shields
If necessary, remove the 14mm and 8 mm T-10 screws holding the IR Shield to the
Main Board and take off the shield.
Note—
If the Main Board has been removed, the IR Shield has been removed already.
4 mm T-6
Troubleshooting, Service and Repair
Repair
14mm T-10
8mm T-10
5-39