Rear Case Disassembly Procedures

Removing the Rear Connector PCB and the Interface Board

Chapter 6 - Disassembly/Reassembly Guide
This section describes the items that may be removed/replaced on the rear case
assembly of the monitor. First separate the front and rear assemblies as described
in the section "Separation of Front and Rear Case Assemblies" on page 45.
The interface PCB is connected at right angles to the rear connector PCB, and is
separable from it.
Step 1. From outside the rear cover, use a 3/16 socket driver to remove the
two standoff fasteners of the RS-232 connector, and also the two
standoff fasteners of the parallel connector.
Standoff
fasteners
Step 2. From inside the rear cover, remove the rear connector PCB and the
interface PCB connected to it.
Rear Case Disassembly Procedures