3.8
Connecting the eluent bottle
Accessories
883 Basic IC plus
The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing . The eluent aspiration tubing is installed on the entry to the high-
pressure pump.
The tubing must be be fed out of the instrument through a suitable open-
ing before the loose end can be connected to the eluent bottle.
For this step, you need the following accessories:
Eluent bottle (6.1608.070)
■
The eluent bottle cap GL 45 accessory set (6.1602.160)
■
This accessory set contains the bottle cap, an M6 tubing nipple, an M8
tubing nipple, two O-rings and an M6 and M8 threaded stopper.
The tubing adapter for aspiration filter accessory set (6.2744.210)
■
This accessory set contains a filter holder, a clamping screw and tubing
weighting.
An aspiration filter (6.2821.090)
■
The adsorber tube (6.1609.000)
■
The SGJ clip (6.2023.020)
■
Connecting the eluent aspiration tubing
1 Installing the eluent bottle cap (6.1602.160)
Start by pushing the M8 tubing nipple onto the loose end of the
■
eluent aspiration tubing, followed by the O-ring.
Push the loose end of the eluent aspiration tubing through the
■
M8 opening of the bottle cap and screw it on for the time being.
Figure 7
Installing the eluent bottle cap
2 Mounting the tubing adapter
Install the parts of the tubing adapter for aspiration filter
(6.2744.210) accessory set:
3 Installation
23
■■■■■■■■