883 Basic IC plus
3 Reinserting the outlet valve into the pump head
Insert the outlet valve into the outlet valve holder (the seal must
be visible) (1).
Screw the outlet valve holder up into the pump head (2) and
tighten it firmly by hand and then retighten it one additional ¼
turn using the adjustable wrench (3).
Tighten the connection capillary to the auxiliary piston back onto
the outlet valve holder.