3.18 Initial start-up
48
■■■■■■■■
3 Starting equilibration
Start the equilibration in MagIC Net: Workplace ▶ Run ▶ Equi-
■
libration ▶ Start HW.
4 Deaerating the high-pressure pump
Push the end of the purge needle (6.2816.040) over the end of
■
the purge capillary on the purge valve.
Insert the syringe (6.2816.020) in the Luer connector of the purge
■
needle.
Open the purge valve using the rotary knob (approx. ½ turn).
■
Switch on the high-pressure pump in MagIC Net.
■
Use the syringe to aspirate eluent until there are no more air bub-
■
bles in the eluent aspiration tubing.
Switch off the high-pressure pump in MagIC Net.
■
Seal the purge valve using the rotary knob.
■
Remove the syringe from the purge needle.
■
Pull the purge needle out of the purge capillary.
■
5 Rinsing the instrument without columns
Rinse the instrument (without columns) with eluent for 10
■
minutes.
■■■■■■■■■■■■■■■■■■■■■■