883 Basic IC plus
Inserting a new filter
1 Inserting a new filter
Use tweezers to carefully place a new filter into the filter screw
■
(24-2) so that it is flat and press it firmly into place with the rear
of the tweezers.
2 Installing the filter screw
Screw the filter screw (24-2) back into the filter housing (24-1)
■
and tighten by hand. Then use two adjustable wrenches
(6.2621.000) to tighten it slightly.
3 Remounting the inline filter
Screw pressure screws back onto the inline filter.
■
Take care to ensure that the flow direction matches that inscribed
on the inline filter.
4 Rinsing the inline filter
Dismantle the guard column (if present) and the separation col-
■
umn and replace with a coupling (6.2744.040).
Rinse the instrument with eluent.
■
Reinsert the columns after 10 minutes.
■