883 Basic IC plus
3 Inserting the pump tubing
Press the tubing cartridge's contact pressure lever down all the
■
way.
Place the pump tubing in the tubing cartridge. Fit the tubing car-
■
tridge between two stoppers. The stoppers must snap into the
corresponding holder of the tubing cartridge.
4 Inserting the tubing cartridge
Hang the tubing cartridge in the mounting bolt and press it in the
■
cartridge holder until you hear the snap-action lever snap in.
Setting the flow rate
The flow rate of the peristaltic pump depends on many factors:
The inner diameter of the pump tubing
■
The rotational speed of the drive
■
The contact pressure of the tubing cartridge
■
NOTICE
Pieces of pump tubing are consumables. The service life of the pump
tubing depends on the contact pressure, among other factors.
Setting the contact pressure correctly
1
Fully loosen the contact pressure lever , i.e. press it all the way
■
down.
In the software, activate the drive of the peristaltic pump with the
■
desired speed.
Raise the contact pressure lever one step at a time until liquid
■
flows.
3 Installation
41
■■■■■■■■