Warm-Up Mode - Siemens ULTRAMAT 23 Operating Instructions Manual

Gas analyzer for ir-absorbing gases and oxygen
Hide thumbs Also See for ULTRAMAT 23:
Table of Contents

Advertisement

5.3

Warm-up Mode

Warm-up mode

Warm-up mode
Time left 02:02 min
AUTOCAL
Time left 01:44 min
Purging with CAL gas
AUTOCAL
Time left 0:44 min
Purging w. meas. gas
Fig. 5-3
Warm-up mode, measuring mode and input mode
53a
Warm-up mode
Time left 02:02 min
53b
AUTOCAL
Time left 01:44 min
Purging with CAL gas
53c
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
The analyzer can be in one of three operating modes:
S
In the warm-up mode,
S
in measuring mode (Section 5.4),
S
or in input mode (Section 5.5).
53a
53b
53c
Meas. mode
403.8
vpm
1875
vpm
936.4
vpm
4.2
%
53d
Immediately following switching-on, the ULTRAMAT 23 tests the
display elements. During this test, all elements light up simultaneously
for approx. five seconds.
The adjacent display subsequently appears with the remaining
warm-up time which is counted down in seconds to 00:00 (minutes :
seconds).
The analyzer initially carries out an AUTOCAL during the warm-up
mode (see Section 4.3.1). The flow of AUTOCAL gas (nitrogen or air)
is displayed in the bottom line, and the remaining time is shown in the
line above this. This calibration cannot be interrupted.
CO
|
NO
|
ENTER
SO
|
2
O
|
2
54
Operation
Input mode
Analyzer status
Calibration
Parameters
Configuration
55
5-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

7mb23357mb23377mb2338

Table of Contents