3.2 Design

• 19" unit with 4 HU for installation
- in swing frame
- in cabinets, with or without slide rails
• Flowmeter for sample gas on front panel (not in version with
pipes);
Option: integrated sample gas pump (standard for bench-top
version)
• Gas connections for sample gas input and output as well as
combustion gas and combustion air, pipe diameter 6 mm or
¼"
• Gas and electrical connections at the rear (portable version:
sample gas inlet at front).
Display and control panel
• Operation according to NAMUR recommendation
• Simple, fast programming and commissioning of analyzer
• Large, backlit LCD for measured values
• Menu-based inputs for programming, test functions and cali-
bration
• Washable membrane keypad
• User help in plain text
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
LED backlit display;
brightness adjustable via menu
One line per component for
measured value, dimension
and component name
Immediate return to
measurement mode
CAL key to start AUTOCAL
with ambient air or N
2
or air without CO
2
Switch internal pump on and off;
pump flowrate adjustable via menu
• Multi-language operation software.
Inputs and outputs
• Three binary inputs for sample gas pump On/Off, triggering of
AUTOCAL and synchronization of several devices
• Eight relay outputs can be freely configured for fault, mainte-
nance request, maintenance switch, limits, range identifica-
tion and external solenoid valves
• Eight additional relay outputs as an option
• Eight additional binary inputs as an option
• Electrically isolated analog outputs.
Communication
• RS 485 present in basic unit (connection from the rear).
Options
• Converter to RS 485/USB
• Converter to TCP/IP Ethernet
• Incorporation in networks via PROFIBUS DP/PA interface
• SIPROM GA software as service and maintenance tool.
Dimension freely selectable
3
(ppm, vpm, mg/m
, %)
Two columns reserved
for status displays
ENTER key to call the
main menu or to save
entered values
↑↓ → Keys for menu control;
increasing/decreasing numerical values
Scroll back in menu or
cancel an input
Technical Description
3-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

7mb23357mb23377mb2338

Table of Contents