Technical Description
Gas path
Key to gas path figures
1
Inlet fo sample gas/ calibration gas
2
Gas outlet
3
Inlet for AUTOCAL/zero gas or
Inlet for sample gas/ adjusting gas (channel 2)
4
Gas outlet (channel 2)
5
Enclosure flushing
6
Inlet atmospheric pressure sensor
7
Inlet/chopper compartment flushing
8
Condensation trap with filter
1
8
13
F
16
2
ULTRAMAT 23, portable, in sheet-steel housing with internal sample gas pump, condensation trap with safety filter on front panel, optional
oxygen measurement
3-6
12
P
10
11
15
15
9
Safety fine filter
10
Solenoid valve
11
Sample gas pump
12
Pressure switch
13
Flowmeter
14
Infrared measuring cell
15
Safety condensate trap
16
Oxygen measuring cell
17
Atmospheric pressure sensor
14
with O
2
without O
2
Operating instructions ULTRAMAT 23 gas analyzer
Gas outlet
2
9
Zero gas
3
(ambient air
or N 2 )
Enclosure
5
flushing
Chopper
7
compartment
flushing