Siemens ULTRAMAT 23 Operating Instructions Manual page 84

Gas analyzer for ir-absorbing gases and oxygen
Hide thumbs Also See for ULTRAMAT 23:
Table of Contents

Advertisement

Operation
Analog output during
function control
Meas. value: hold
51011b
5-34
Table 5-2
Analog current output for start-of-scale values
selectable
0 - 20 mA
2 - 20 mA
4 - 20 mA
NAMUR - 20 mA
Table 5-3
Analog output current for downward limiting of measured
value
selectable
0 - 20 mA
2 - 20 mA
4 - 20 mA
NAMUR - 20 mA
Table 5-4
Analog output current for upward limiting of measured value
selectable
0 - 20 mA
2 - 20 mA
4 - 20 mA
NAMUR - 20 mA
1) if measuring range 2 is set to the maximum possible value, the limiting
value of range 2 ... 20 mA is at 20.9 mA and of range 4 ... 20 mA at
20.8 mA.
S
Analog output during FCTRL (analog output during function
control)
The following measured value outputs are possible with the
AUTOCAL procedure, in warm-up mode and in the uncoded state:
Hold: The value measured directly prior to commencement is
output unchanged. This also applies to the limits which are
output (see section 5.9.2)
Actual: The measured value is continuously updated.
Zero: see following table:
Analog current (neg. on)
0 mA
2 mA
4 mA
4 mA
Analog current (neg. on)
0.0 mA
0.0 mA
0.0 mA
3.8 mA
Analog current (neg. on)
21.0 mA
21.0 mA
21.0 mA
20.5 mA
Operating instructions ULTRAMAT 23 gas analyzer
Analog current (neg. off)
0 mA
2 mA
4 mA
4 mA
Analog current (neg. off)
0.0 mA
2.0 mA
4.0 mA
4.0 mA
Analog current (neg. off)
21.0 mA
21.0 mA
21.0 mA
20.5 mA
C79000-B5276-C216-03