Fault Messages - Siemens ULTRAMAT 23 Operating Instructions Manual

Gas analyzer for ir-absorbing gases and oxygen
Hide thumbs Also See for ULTRAMAT 23:
Table of Contents

Advertisement

6.1.2

Fault Messages

Mains voltage beyond
tolerance
press ENTER to clear
next message with ->
Table 6-1

Fault messages

Fault message
Meas. value channel 1 beyond
tolerance
Meas.value display: *****
Meas. value channel 2 beyond
tolerance
Meas.value display: *****
Meas. value channel 3 beyond
tolerance
Meas.value display: *****
Meas. value O
beyond tolerance
2
Meas. value display: *****
Mains voltage beyond tolerance
Temp. of analyzer beyond
tolerance
Pressure of amb. air beyond
tolerance
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
Fault messages are references to certain changes in the analyzer
which influence its ability to measure correctly.
If the analyzer is in measuring mode, the occurrence of a fault is
indicated by an "F" at the right edge of the display.
Fault messages are logged and can be called in input mode using the
menu path "Analyzer status - Status - Logbook/faults" (see Section
5.7.1.1). The corresponding message texts are stored as alphanumeric
text in the logbook. You can delete the messages by pressing the
ENTER key. However, they appear again if the cause has not been
eliminated.
Table 6-1 provides a summary of fault messages, their causes, and
measures to eliminate the faults. If a remedy is not listed, you must
contact the servicing department when this fault message occurs.
Possible causes
Analyzer section of first
component is faulty
Analyzer section of second
component is faulty
Analyzer section of third
component is faulty
O
sensor faulty or old
2
Motherboard faulty
Mains voltage varies
Power supply unit faulty
Ambient temperature too high or
too low
Temperature sensor of a detector
faulty
Pressure sensor faulty
Maintenance
Remedy
Replace the O
sensor as
2
described in Section 6.2.1.
Correct such that the mains
voltage is within the tolerance
values permissible for the
analyzer.
Provide sufficient ventilation or air
conditioning.
6-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

7mb23357mb23377mb2338

Table of Contents