Siemens ULTRAMAT 23 Operating Instructions Manual page 85

Gas analyzer for ir-absorbing gases and oxygen
Hide thumbs Also See for ULTRAMAT 23:
Table of Contents

Advertisement

Analog output
during fault
Meas. value: hold
51011c
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
selectable
0 - 20 mA
2 - 20 mA
4 - 20 mA
NAMUR - 20 mA
21 mA: see following table:
selectable
0 - 20 mA
2 - 20 mA
4 - 20 mA
NAMUR - 20 mA
1) If measuring range 2 is set to the maximum possible value, the
limiting value of range 2 ... 20 mA is at 20.9 mA and of
range 4 ... 20 mA at 20.8 mA.
S
Output during faults (analog output during a fault)
It is possible for you to define the type of measured value output
during a fault. The following are possible:
Hold: The last value measured prior to the fault is output for the
duration of the fault.
Actual: The measured value is continuously updated during the
fault.
Zero: see following table:
selectable
0 - 20 mA
2 - 20 mA
4 - 20 mA
NAMUR - 20 mA
21 mA: see following table:
selectable
0 - 20 mA
2 - 20 mA
4 - 20 mA
NAMUR - 20 mA
1) If measuring range 2 is set to the maximum possible value, the
limiting value of range 2 ... 20 mA is at 20.9 mA and of
range 4 ... 20 mA at 20.8 mA.
Analog current (neg. on)
0 mA
2 mA
4 mA
3 mA
Analog current (neg. on)
21.0 mA
21.0 mA
21.0 mA
21.5 mA
Analog current (neg. on)
0 mA
2 mA
4 mA
3 mA
Analog current (neg. on)
21.0 mA
21.0 mA
21.0 mA
21.5 mA
Operation
Analog current (neg. off)
0 mA
2 mA
4 mA
3 mA
Analog current (neg. off)
21.0 mA
1)
21.0 mA
1)
21.0 mA
21.5 mA
Analog current (neg. off)
0 mA
2 mA
4 mA
3 mA
Analog current (neg. off)
21.0 mA
1)
21.0 mA
1)
21.0 mA
21.5 mA
5-35