Shut-down
7.1
Measures for Repair or Changing of Location
19" rack version
Benchtop version
7.2
Measures for Scrapping the Analyzer
19" rack version
7-2
It may be necessary to shut down the ULTRAMAT 23 for the following
reasons:
S
Repair
S
New location of use
S
Scrapping
If the ULTRAMAT 23 is shut down for repair or for changing the
location of use, proceed as follows:
S
Make sure that gas is no longer flowing through the analyzer. If external
pumps are present, switch all of them off.
S
Switch the analyzer off.
S
Disconnect the power plug.
S
Disconnect all hose connections from the rear of the analyzer.
S
With pipe versions, unscrew all pipes.
S
Make sure that gas is no longer flowing through the analyzer. If external
pumps are present, switch all of them off.
S
Switch the analyzer off.
S
Disconnect the power plug.
S
Empty the condensation trap (see Section 6.2.4).
S
Disconnect the supply hose from the condensation trap.
S
Disconnect all hose connections from the rear of the analyzer
If the ULTRAMAT 23 is to be scrapped, shut down as follows:
S
Make sure that gas is no longer flowing through the analyzer. If
external pumps are present, switch all of them off.
S
Switch the analyzer off.
S
Disconnect the power plug.
S
Disconnect all hose connections from the rear of the analyzer.
S
With pipe versions, unscrew all pipes.
S
In the case of analyzers with an electrochemical oxygen sensor,
remove this from the analyzer (see Section 6.2.1).
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03