Siemens ULTRAMAT 23 Operating Instructions Manual page 93

Gas analyzer for ir-absorbing gases and oxygen
Hide thumbs Also See for ULTRAMAT 23:
Table of Contents

Advertisement

Address :
125
Ident number
: 1
PB relay: OFF
dIAGNOSIS
51023b
INTERF: NNNNN
const.Interfer: OFF
C.I: +X.XXXXe+X
51023c
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
S
PROFIBUS parameters
If the analyzer has a PROFIBUS-PA/DP option module, you can set
here the station address as a decimal number as well as further
PROFIBUS parameters.
Notice
For further details on these PROFIBUS parameters, please refer to the
PROFIBUS-PA/DP interface description (A5E00054148).
S
ELAN const. Interf.
The adjacent display appears following selection of this parameter.
This function is used to switch on or off the correction of the
influence of a constant interference. If the correction function is on,
the set correction value is added to the measured value of the
corresponding canal.
Example:
A constant interference of +10 ppm is detected. You enter -10 ppm
as correction value. Thus the measured value will be modified from
+10 ppm to 0 ppm.