Analyzer Status: Status: Logbook/Faults

Mains voltage
beyond tolerance
Press ENTER to clear
Next message with ->
5711
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
Here you can view all analyzer data. The menu sequence in Fig. 5-4
shows all submenus which can be accessed from the menu "Analyzer
status". The arrows lead by one menu item to the next lower
menu/dialog level which is called by this menu item.
In this menu you can call the status messages of the ULTRAMAT 23
via further menu items.
This dialog displays a logbook of all recorded fault messages. Each
type of fault message only appears once in the logbook and is output in
alphanumeric text (see Chapter 6 for list of possible fault messages).
Viewing the logbook is protected by code level 1.
You can:
S
Display all recorded faults in succession using the → key
S
Delete the currently displayed fault message using the ENTER key,
whereupon the next fault message appears, if present. You should
delete those fault messages whose causes have been eliminated.
Notice
By deleting the fault message you do not eliminate the cause of the
fault (see Chapter 6)!
S
A corresponding text is output when all stored fault messages have
been displayed. Terminate display of the logbook using the → key.
Notice
If the analyzer is in measuring mode, you can recognize the
occurrence of a fault in that an "F" appears at the right edge. A "!" at
the right edge signals that a fault has been logged which is no longer
currently present.