Using the Pump

Solvent Information

Solvent Information

Observe the following recommendations on the use of solvents.
• Follow recommendations for avoiding the growth of algae, see
• Small particles can permanently block capillaries and valves. Therefore,
• Avoid or minimize the use of solvents that may corrode parts in the flow path.
Materials in Flow Path
Following materials are used in the flow path of this module:
Table 6
Part
Degasser chamber
SSV
Active inlet valve
Outlet valve
Adapter
Pump head (body)
Pistons
Piston seals/wash seals
Pressure sensor
Purge valve
Damping unit
Capillaries/fittings
Tubings
48
Growth in HPLC Systems"
always filter solvents through 0.22 μm filters.
Consider specifications for the pH range given for different materials like flow
cells, valve materials etc. and recommendations in subsequent sections.
Materials in flow path
Materials
TFE/PDD Copolymer, PFA (internal tubings), PEEK (inlets), FEP
(tubings), ETFE (fittings)
PEEK, FFKM
SST, sapphire, ruby, ceramic, PTFE
SST, gold, ruby, ZrO
SST, gold
SST
Sapphire
PTFE, SST (reversed phase) or UHMW-PE, SST (normal phase)
SST
SST, gold, PTFE, ceramic
SST, gold
SST
PTFE
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
on page 54
-based ceramic, tantalum