Solvent Information

Introduction
Observe the following recommendations on the use of solvents.
• Follow recommendations for avoiding the growth of algae, see
• Small particles can permanently block capillaries and valves. Therefore,
• Avoid or minimize the use of solvents that may corrode parts in the flow
1290 Infinity Quaternary Pump User Manual
Growth in HPLC Systems"
always filter solvents through 0.22 μm filters.
path. Consider specifications for the pH range given for different materials
like flow cells, valve materials etc. and recommendations in subsequent
sections.
on page 89.
Using the Module

Solvent Information

"Algae
4
77