OXYGEN SUPPLY

WARNING
WARNING

HIGH PRESSURE OXYGEN SUPPLY

WARNING
CAUTION
7.
Charge the battery by following the instructions in Full Recharge Procedure,
page 36.
The iVent
201 can use medical-grade oxygen from either
TM
A high-pressure source such as a central supply system or cylinder at 40-75
psi (2.8-5.1 bar).
The exact oxygen pressure which the iVent
connected to medical grade oxygen depends on the version you have.
Check the back label (Shown in Figure 11) for the pressure levels of your
unit.
A low-pressure oxygen source -- oxygen concentrator or flow meter device -
using the optional Low Pressure Oxygen Enrichment System (M1161007,
M1162020, M1162022)
When connecting the iVent
type of oxygen source is selected in Advanced Settings. [Main
Screen/Menu/Advanced Settings/Oxygen Supply (Pressure)] (See page 94). The
factory default is High pressure.
The O2 system must be calibrated every 6 months
We recommend calibration every 3 months to ensure the O2 system integrity. For
instructions for performing the O2 calibration see O2 Calibration Page 177.
If using pressurized oxygen, connect the oxygen supply to the oxygen O2 inlet
connector at the back of the ventilator (Figure 11).
To prevent explosion hazard always ensure the oxygen connector is free
from oil.
Verify that the oxygen supply is at the correct pressure before connecting to
the O2inlet.
When the unit is ventilating the measured concentration of delivered oxygen can
be viewed on the Alarm Settings screen (See page 145).
201 to an oxygen supply, ensure that the correct
TM
38
201 requires when
TM